Products

low profile industrial icp® accel, 100 mv/g, 0.5 to 10k hz, side exit, 10-ft armored jacketed cable & swiveler base, single point iso 17025 accredited calibration

Spindler® Industrial ICP® Accelerometer

Model 607A61

Retrieving Price...

(Pricing valid only in the U.S.)

Units in Stock Lead Times
16 Lead Time for 1-5 is 3 Wks.
Lead Time for 6-10 is 4 Wks.
Configure
  • CE
  • IP

Low profile industrial ICP® accel, 100 mV/g, 0.5 to 10k Hz, side exit, 10-ft armored jacketed cable & swiveler base, single point ISO 17025 accredited calibration

US patent number 6.435.902
  • Sensitivity: (±15%)100 mV/g (10.2 mV/(m/s²))
  • Frequency Range: (±3dB)30 to 600000 cpm (0.5 to 10000 Hz)
  • Sensing Element: Ceramic
  • Measurement Range: ±50 g (±490 m/s²)
  • Weight: 1.1 oz (31 gm)

Additional configurations:

Specification Notes


[1] Conversion Factor 1g = 9.81 m/s².
[2] Zero-based, least-squares, straight line method.
[3] Typical.
[4] 1/4-28 has no equivalent in S.I. units.
[5] Stud torque must exceed sensor hex nut torque to ensure proper dismantling.
[6] Measured with mounting stud.
[7] Stainless steel armor jacket over twisted shielded pair.
[8] 1/8" hex Allen key required for English version, 3mm hex Allen key required for metric version.
[9] See PCB Declaration of Conformance PS023 or PS060 for details.

English: SI:
Performance
Sensitivity (±15 %) 100 mV/g 10.2 mV/(m/s²) [1]
Measurement Range ±50 g ±490 m/s²
Frequency Range (±3 dB) 30 to 600000 cpm 0.5 to 10000 Hz
Resonant Frequency 1500 kcpm 25 kHz [3]
Broadband Resolution (1) 350 µg 3434 µm/sec2 [3]
Non-Linearity ±1 % ±1 % [2]
Transverse Sensitivity ≤7 % ≤7 %
Environmental
Overload Limit (Shock) 5000 g pk 49050 m/s² pk
Temperature Range -65 to +250 °F -54 to +121 °C
Temperature Response See Graph %/°F See Graph %/°F [3]
Enclosure Rating IP67 IP67
Electrical
Settling Time (within 1% of bias) ≤2 sec ≤2 sec
Discharge Time Constant ≥0.3 sec ≥0.3 sec
Excitation Voltage 18 to 28 VDC 18 to 28 VDC
Constant Current Excitation 2 to 20 mA 2 to 20 mA
Output Impedance <150 Ohm <150 Ohm
Output Bias Voltage 8 to 12 VDC 8 to 12 VDC
Spectral Noise (10 Hz) 8 µg/√Hz 78.5 (µm/sec2)/√Hz [3]
Spectral Noise (100 Hz) 5 µg/√Hz 49.1 (µm/sec2)/√Hz [3]
Spectral Noise (1 kHz) 4 µg/√Hz 39.2 (µm/sec2)/√Hz [3]
Electrical Isolation (Case) >100000000 Ohm >100000000 Ohm
Physical
Size - Hex 9/16 in 14 mm
Size - Height 1.0 in 25.4 mm
Weight (without cable) 1.1 oz 31 gm [6]
Mounting Stud Stud
Mounting Thread 1/4-28 Male 1/4-28 Male [4]
Mounting Torque (Stud) 3 to 4 ft-lb 4.1 to 5.4 Nm [5] [8]
Mounting Torque (hex nut) 2 to 3 ft-lb 2.7 to 4.1 Nm
Sensing Element Ceramic Ceramic
Sensing Geometry Shear Shear
Housing Material Stainless Steel Stainless Steel
Sealing Welded Hermetic Welded Hermetic
Electrical Connector Integral Armored Cable Integral Armored Cable
Electrical Connection Position Side Side
Cable Length 10 ft 3.0 m
Cable Type Polyurethane Polyurethane [7]

* All specifications are at room temperature unless otherwise specified.

TempChart

Accessories


Supplied
ICS-2 Calibration of industrial accel., single-point, 100 Hz reference sensitivity 1
080A156 Swivel mounting stud, 1/2-20 to 1/4-28 stud with 1/8" hex socket for installation (for models 607A11 & 607A61) 1

Optional Versions


(EX) Hazardous Area Approval- contact factory for specific approvals

English SI
Hazardous Area Approval EEx nL IIC 1T4, -40 = Ta = 121°C, II 3 G
Hazardous Area Approval EEx nL IIC T4, -40 ° C ≤ Ta ≤ 121 ° C, II 1 G
Hazardous Area Approval Cl I, Div I, Groups A, B, C, D; Cl II, Div I, Groups E, F, G; Cl III, Div I
Hazardous Area Approval Cl I, Div 2, Groups A, B, C, D; ExnL IIC T4, AExnA IIC T4
Hazardous Area Approval Ex ia IIC T4, AExia IIC, T4
(M) Metric Mount
Accessories:
Supplied M080A159A Accessory for IMI Sensors Product 1

(TO) Temperature Output

English SI
Temperature Output Range +36 to +250 °F +2 to +121 °C
Temperature Scale Factor 5.56 mV/°F + 32 +10 mV/°C
Electrical Connector Integral Armored Cable
Electrical Connections Acceleration Output
Electrical Connections Ground
Electrical Connections Temperature Output