tourmaline icp® underwater blast sensor, 7.6 length, 5000 psi, 1 mv/psi

ICP® Pressure Sensor

Model 138A05

Retrieving Price...

(Pricing valid only in the U.S.)

Units in Stock Lead Times
More on the way Call for Lead Time
  • CE
  • IP

Tourmaline ICP® underwater blast sensor, 7.6" length, 5000 psi, 1 mV/psi

Photo shown with special W option attached underwater cable. Standard model supplied with 10-32 jack.
  • Measurement Range: 5000 psi (34475 kPa)
  • Sensitivity: (±15%)1.0 mV/psi (0.15 mV/kPa)
  • Low Frequency Response: (-5%)1.7 Hz
  • Resonant Frequency: >=1000 kHz
  • Electrical Connector: 10-32 Coaxial Jack
  • Weight: 0.75 oz (21.0 gm)

Specification Notes


[1] Zero-based, least-squares, straight line method.
[2] Typical.
[3] For +10 volt output, minimum 24 VDC supply voltage required. Negative 10 volt output may be limited by output bias.
[4] See PCB Declaration of Conformance PS023 for details.

English: SI:
Performance
Measurement Range (for ±5V output) 5000 psi 34475 kPa
Useful Overrange (for ± 10V output) 10000 psi 68950 kPa [3]
Sensitivity (±15 %) 1.0 mV/psi 0.15 mV/kPa
Maximum Pressure 50000 psi 344750 kPa
Resolution 0.1 psi 0.7 kPa [2]
Resonant Frequency ≥1000 kHz ≥1000 kHz
Rise Time (Reflected) ≤1.5 µ sec ≤1.5 µ sec
Low Frequency Response (-5 %) 1.7 Hz 1.7 Hz
Non-Linearity ≤2.0 % FS ≤2.0 % FS [1]
Environmental
Temperature Range (Operating) -10 to +100 °F -23 to +37.8 °C
Maximum Shock 20000 g pk 196140 m/s² pk
Electrical
Output Polarity (Positive Pressure) Positive Positive
Discharge Time Constant (at room temp) ≥0.3 sec ≥0.3 sec
Excitation Voltage 20 to 30 VDC 20 to 30 VDC
Constant Current Excitation 2 to 20 mA 2 to 20 mA
Output Impedance ≤100 Ohm ≤100 Ohm
Output Bias Voltage 8 to 14 VDC 8 to 14 VDC
Physical
Sensing Element Tourmaline Tourmaline
Housing Material Stainless Steel Stainless Steel
Electrical Connector 10-32 Coaxial Jack 10-32 Coaxial Jack
Weight 0.75 oz 21.0 gm

* All specifications are at room temperature unless otherwise specified.